Differential Probes

Comparison Chart

Typ Attenuation Ratio Bandwidth DC-Accuracy max. Input Voltage Differential Range (DC + Peak AC)
Series 7000
TT-SI-7002 x20/x50/x200 25MHz 2% ±70V / ±175V / ±700V
TT-SI 7005 x1/x5/x10 60MHz 2% ±3.5V / ±18V / ±35V
Series 8000
TT-SI 8071 x10/x100 70MHz 2% ±70V / ±700V
TT-SI 8050 x50/x500 70MHz 2% ±150V / ±1500V
TT-SI 8051 x50/x500 100MHz 2% ±150V / ±1500V
TT-SI 8052 x50/x500 200MHz 2% ±150V / ±1500V
TT-SI 8110 x100/x1000 100MHz 2% ±280V / ±2800V
TT-SI 8010A x100/x1000 70MHz 2% ±700V / ±7000V
TT-SI 8010B x100/x1000 100MHz 2% ±700V / ±7000V
Series 9000
TT-SX 9001 x10/x100 30MHz 2% ±70V / ±700V
TT-SX 9002 x20/x200 30MHz 2% ±140V / ±1400V
Series SX
TT-SX 200 x10 200MHz 2% ±60V
Download Comparison Chart

Differential Probes Series TT-SI 7000

TT-SI 7002

Supplied Accessories
Bandwidth25 MHz @ x50 and x200
15 MHz @ x20
Attenuation Ratiox20 / x50 / x200
DC-Accuracy±2 %
Input Impedance
Differential Mode
4 MΩ, 1.2 pF
Input Impedance2 MΩ, 2.3 pF
Rise Time14 ns @ x50 and x200
23 ns @ x20
Output Voltage Swing±8 V (using 1 MΩ oscilloscope input)
Offset±5 mV
Noise2 mVrms
Source Impedance50 Ω
CMRR (typical)80 dB @ 60 Hz, 60 dB @ 100 Hz, 50 dB @ 1 MHz
Max. Input Voltagex20 ±70 V (DC + peak AC)
x50 ±175 V (DC + peak AC)
x200 ±700 V (DC + peak AC)
Input Voltage
Absolute max. Voltage
600 Vrms
Power requirements9 V mains adapter (included)
Operation Temperature0 °C to 50 °C
Storage Temperature-30 °C to 70 °C
Dimensions (L x W x H)195 mm x 55 mm x 30 mm
CableBNC Cable 100 cm
Input Leads 55 cm
Weight250 g
IEC61010-031: 2015CAT III Pollution Degree 2
Data Sheet
Manual
Order Number15175

Only connect to a grounded measuring device!


TT-SI 7005

Supplied Accessories
Bandwidth60 MHz @ x10
50 MHz @ x5
15 MHz @ x1
Attenuation Ratiox1 / x5 / x10
DC-Accuracy±2 %
Input Impedance
Differential Mode
1 MΩ, 2.5 pF
Input Impedance0,5 MΩ, 6 pF
Rise Time5.8 ns @ x10
7 ns @ x5
23 ns @ x1
Output Voltage Swing±4 V (using 1 MΩ oscilloscope input)
Offset±5 mV
Noise2 mVrms
Source Impedance50 Ω
CMRR (typical)80 dB @ 60 Hz, 60 dB @ 100 Hz, 50 dB @ 1 MHz
Max. Input Voltagex1 ±3.5 V (DC + peak AC)
x5 ±18 V (DC + peak AC)
x10 ±35 V (DC + peak AC)
Input Voltage
Absolute max. Voltage
35 V
Power requirements9 V mains adapter (included)
Operation Temperature0 °C to 50 °C
Storage Temperature-30 °C to 70 °C
Dimensions (L x W x H)195 mm x 55 mm x 30 mm
CableBNC Cable 100 cm
Input Leads 55 cm
Weight250 g
IEC61010-031: 2015CAT III Pollution Degree 2
Data Sheet
Manual
Order Number15176

Only connect to a grounded measuring device!

Differential Probes Series TT-SI 8000

TT-SI 8071 and TT-SI 8110

TT-SI 8071 TT-SI 8110
Order Number 15401 15410
Bandwidth 70MHz 100MHz
Attenuation x10 / x100 x100 / x1000
DC-Accuracy ±2%
Max. Input Voltage
Differential Mode
±70V (DC + AC peak) @ x10
±700V (DC + AC peak) @ x100
±280V (DC + AC peak) @ x100
±2800V (DC + AC peak) @ x1000
Max. Input Voltage
Common Mode
±700V (DC + AC peak) ±2800V (DC + AC peak)
Max. Input Voltage
(each side to ground)
600Vrms CAT III
1000Vrms CAT II
Safety Standard IEC1010-031:2015
Data Sheet
Specifications
Manual
Supplied Accessories 2 x Test Leads with insulated 4mm Banana Plugs (red & black)
2 x Grabber (red & black), 2 x Pincer (red & black)
2 x Alligator Clips (red & black), 1 x Insulated BNC Cable 100cm
1 x USB power lead, 1 x Mains Adapter 220V EU Version

Only connect to a grounded measuring device!


TT-SI 8050, TT-SI 8051 and TT-SI 8052

TT-SI 8050 TT-SI 8051 TT-SI 8052
Order Number 15402 15403 15404
Bandwidth 70MHz 100MHz 200MHz
Attenuation x50 / x500
DC-Accurac ±2%
Max. Input Voltage
Differential Mode
±150V (DC + AC peak) @ x50
±1500V (DC + AC peak) @ x500
Max. Input Voltage
Common Mode
±1500V (DC + AC peak)
Max. Input Voltage
(each side to ground)
600Vrms CAT III
1000Vrms CAT II
Safety Standard IEC1010-031:2015
Data Sheet
Specifications
Manual
Supplied Accessories 2 x Test Leads with insulated 4mm Banana Plugs (red & black)
2 x Grabber (red & black), 2 x Pincer (red & black)
2 x Alligator Clips (red & black), 1 x Insulated BNC Cable 100cm
11 x USB power lead, 1 x Mains Adapter 220V EU Version

Only connect to a grounded measuring device!


TT-SI 8010A and TT-SI 8010B

TT-SI 8010A TT-SI 8010B
Order Number 15411 15412
Bandwidth 70MHz 100MHz
Attenuation x100 / x1000
DC-Accuracy ±2%
Max. Input Voltage
Differential Mode
±700V (DC + AC peak) @ x10
±7000V (DC + AC peak) @ x100
Max. Input Voltage
Common Mode
±7000V (DC + AC peak)
Max. Input Voltage
(each side to ground)
1000Vrms CAT III
2300Vrms
Safety Standard IEC1010-031:2015
Data Sheet
Specifications
Manual
Supplied Accessories 2 x Test Leads with insulated 4mm Banana Plugs (red & black)
2 x Grabber (red & black), 2 x Pincer (red & black)
2 x Alligator Clips (red & black), 1 x Insulated BNC Cable 100cm
1 x USB power lead, 1 x Mains Adapter 220V EU Version

Only connect to a grounded measuring device!


Differential Probes Series TT-SX 9000 - NEW

TT-SX 9001 and TT-SX 9002

TT-SX 9001 TT-SX 9002
Order Number 15601 15602
Bandwidth 30MHz
Attenuation x10 / x100 x20 / x200
DC-Accuracy ±2%
Risetime 14ns
Max. Input Voltage
Differential Mode
±70V @ x10
±700V @ x100
±140V @ x20
±1400V @ x200
Max. Input Voltage
Common Mode
±70V @ x10
±700V @ x100
±140V @ x20
±1400V @ x200
Max. Input Voltage
(each side to ground)
1000V CAT III
Safety Standard IEC1010-031:2015 CAT III
Data Sheet
Manual
Supplied Accessories 1 x TT-SX GR 1 Probe Hook, Pair (red & black)
1 x TT-SX USB Power Cord, 100cm

Only connect to a grounded measuring device!


Differential Probe Series TT-SX 200 - NEW

TT-SX 200

TT-SX 200
Order Number 15635
Bandwidth 200MHz
Attenuation 10:1
DC-Accuracy ±1%
Max. Input Voltage
Differential Mode
±20V (DC + AC peak)
Max. Input Voltage
Common Mode
±60V (DC + AC peak)
Max. Input Voltage
(each side to ground)
60V CAT III
Probe Load requirement 50 Ohm
Power requirements USB power adapter (included)
USB cable (included)
Datasheet
Manual
Supplied Accessories 1 x 50 Ohm Feedthrough Termination, 1 x Hardcase, 2 x Grabber (red & black)
2 x Pincer (red & black), 2 x Alligator Clips (red & black)
1 x Insulated BNC Cable 100cm, 1 x USB power lead, 1 x USB Power Adapter 220V EU Version

Only connect to a grounded measuring device!


Accessories Differential Probes TT-SI 7000 Series

TT-SI NT7 - Power Supply

Order number15163
Cable Length190 cm

TT-SI GR2 - Grabber Red and Black

Order number15104

TT-SI TL7

Order number15164

TT-SI BN7

Order number15165

All specifications and characteristics are subject to change without notice.